1: Anesth Analg. 2009 Feb;108(2):518-20.
LinkOut
The effects of multiple infusion line extensions on
occlusion alarm function of an infusion pump.
Deckert D,
Buerkle C,
Neurauter A,
Hamm P,
Lindner KH,
Wenzel V.
Department of Anesthesiology and Critical Care
Medicine, Innsbruck Medical University, Anichstrasse 35, 6020 Innsbruck, Austria.
diana.deckert@uki.at
BACKGROUND: For anesthesia or conscious sedation of
patients undergoing diagnostic or therapeutic procedures in computed tomography or magnetic resonance imaging
scans, an extension of infusion lines for continuous drug delivery of anesthetics or vasopressors is often
necessary. In this study, we tried to determine if the length of the infusion line influenced the time until
an alarm sounded after occlusion at the end of the infusion line. METHODS: We connected 2 infusion pump
systems of the same model with 1, 2 or 3 infusion lines in series or with a spiral nonkinking low compliance
infusion line, and started the infusion for 60 s. The end of the infusion line was then occluded by turning a
stopcock to occlude the fluid flow. A pressure sensor was connected to the infusion line to record the actual
pressure change in the line. The time until the pressure occlusion alarm sounded was measured 5 consecutive
times at flow rates of 5, 20, and 50 mL/h. RESULTS: When using a single infusion line, pressure occlusion
alarms were triggered after 2.4 +/- 0.1 min for infusion pump 1 and 2.6 +/- 0.2 min for infusion pump 2 at 50
mL/h, after 6.6 +/- 0.4 min and 5.6 +/- 0.5 min at 20 mL/h, and after 23.0 +/- 2.8 min and 20.9 +/- 3.6 min
at 5 mL/h, respectively. When adding a second infusion line, a pressure occlusion alarm was triggered after
27.1 +/- 1.8 min for infusion pump 1 (P = 0.1) and after 29.2 +/- 1.4 min for infusion pump 2 (P = 0.07) at 5
mL/h. With 3 infusion lines, the pressure occlusion alarm of infusion pumps 1 and 2 were significantly
prolonged when compared with 1 infusion line and were released at 31.6 +/- 3.0 min (P = 0.01) and 35.1 +/-
1.1 min (P = 0.001) at 5 mL/h, respectively. The pressure level triggering an alarm ranged in both infusion
pumps between about 900 and 1100 Mbar. CONCLUSIONS: When simulating low flow rate infusions (5 mL/h) as for
vasopressor support, occlusion alarm time was critically prolonged, especially with an increased length of
infusion lines.
PMID: 19151281 [PubMed - indexed for
MEDLINE]
Related Articles
·
Infusion technology: a cause for alarm.
[Paediatr Anaesth. 2002]
·
[The effect of air within the infusion syringe on drug delivery of syringe pump infusion
systems] .
[Anaesthesist. 2000]
·
[Causes of failure and dangers in the use of motor driven infusion pumps. Accidental closure of the infusion system] [Anaesthesist.
1998]
·
Syringe pumps take too long to give occlusion alarm.
[Paediatr Anaesth. 2005]
|